The Beer–Lambert law links how much light a solution absorbs to how much of the absorbing substance it contains. For a single absorbing species at a fixed wavelength, the absorbance is proportional to the molar absorption coefficient , the path length of the cuvette and the molar concentration : . That simple proportionality is what makes UV–visible spectrophotometry the everyday tool for measuring DNA and protein concentrations, following enzyme kinetics with NADH, or checking the colour of a dye bath.
This calculator solves the law for any one of the four quantities:
- concentration from a measured absorbance (the most common use);
- absorbance expected for a known concentration, to plan a dilution that stays in the instrument's best range;
- molar absorption coefficient from a standard of known concentration;
- path length, for micro-volume instruments or flow cells.
You can enter the measurement either as absorbance or as percent transmittance; the calculator converts between them with and always shows the share of light transmitted and absorbed. Concentrations can be in mol/L, mmol/L, µmol/L or nmol/L. To turn a molar result into grams per litre, use the molarity calculator.
How to use this calculator
- Choose what to solve for: concentration, absorbance, ε or path length.
- Enter the measurement as absorbance or as transmittance in %.
- Enter ε in L·mol⁻¹·cm⁻¹ at the wavelength you measured. Values from the literature are wavelength- and solvent-specific.
- Enter the path length (1 cm for a standard cuvette) and the concentration with its unit, unless one of them is the unknown.
- Read the result, the transmittance and the steps. A warning appears when the absorbance lies outside the reliable range.
The formula
The Beer–Lambert law, with decadic absorbance:
where is the transmittance (fraction of the incident intensity that passes through), the molar absorption coefficient in L·mol⁻¹·cm⁻¹, the path length in cm and the amount concentration in mol/L. Rearranged:
Absorbances add up: with several absorbing species, , provided they do not interact. Always subtract a blank (the same cuvette filled with solvent) so that only reflects the analyte.
Worked example
NADH assay. A sample read at 340 nm in a 1 cm cuvette gives . With L·mol⁻¹·cm⁻¹:
so the NADH concentration is 50 µmol/L. That absorbance corresponds to of the light transmitted.
Measuring ε. A 25 µmol/L standard of a dye gives in a 1 cm cell, so 24,800 L·mol⁻¹·cm⁻¹.
Planning a dilution. A 120 µmol/L NADH solution in a 0.5 cm cell would read — comfortably within the linear range.
Transmittance. 10 % transmittance is exactly ; 1 % is .
Where the law stops working
Beer–Lambert is a limiting law: it holds for dilute solutions, monochromatic light and a single, non-interacting absorber. Above roughly – (less than 3–10 % of the light getting through) most instruments lose linearity, because a small amount of stray light becomes comparable to the transmitted beam, and molecules at high concentration interact and change their absorption. Very low absorbances (below about 0.1) suffer from noise and from small differences between cuvettes.
Chemical equilibria also cause deviations: an indicator whose acid and base forms absorb differently, or a dye that dimerises as it gets more concentrated, will not give a straight calibration line. Scattering by turbid samples, bubbles or fingerprints adds apparent absorbance. The safest practice is a calibration curve with standards spanning your samples, and dilution of anything that reads above the linear range (see the dilution factor calculator).
Typical molar absorption coefficients
| Substance | Wavelength | ε (L·mol⁻¹·cm⁻¹) |
|---|---|---|
| NADH, NADPH | 340 nm | 6,220 |
| p-Nitrophenolate (alkaline) | 405 nm | about 18,000 |
| Tryptophan | 280 nm | about 5,500 |
| Tyrosine | 280 nm | about 1,490 |
Values depend on solvent, pH and temperature; use the figure stated in your assay protocol whenever one exists. For proteins, ε at 280 nm is often estimated from the numbers of tryptophan and tyrosine residues.
Frequently asked questions
How do I find concentration from absorbance?
Divide the absorbance by the product of the molar absorption coefficient and the path length: c = A / (ε·l). With A = 0.311, ε = 6220 L·mol⁻¹·cm⁻¹ and l = 1 cm, c = 5.0 × 10⁻⁵ mol/L, i.e. 50 µmol/L.
How do I convert absorbance to transmittance?
Use T = 10^(−A) and multiply by 100 for a percentage. A = 0 means 100 % transmittance, A = 1 means 10 %, A = 2 means 1 %. In the other direction, A = −log₁₀(T) with T as a fraction.
What are the units of the molar absorption coefficient?
With the path length in centimetres and the concentration in mol/L, ε is in L·mol⁻¹·cm⁻¹ (often written M⁻¹·cm⁻¹). Absorbance itself has no unit. If you work in mass concentration (g/L), the matching coefficient is in L·g⁻¹·cm⁻¹.
What absorbance range is reliable?
For most bench spectrophotometers, roughly 0.1 to 1.0, and up to about 1.5 for good instruments. Outside that range, dilute or concentrate the sample, or change the path length, so the reading falls in the linear part of the calibration.
Is the Beer–Lambert law the same as Beer’s law?
Yes. Lambert described the dependence on path length and Beer the dependence on concentration; the combined relation A = εlc is called the Beer–Lambert law, the Beer–Lambert–Bouguer law or simply Beer’s law.
Sources
- IUPAC Compendium of Chemical Terminology (Gold Book): Beer–Lambert law
- IUPAC Compendium of Chemical Terminology (Gold Book): absorbance
- Harvey D., Analytical Chemistry 2.1, §10.2 Spectroscopy Based on Absorption (LibreTexts)